Accuthermo Technology Corp

Loading...

  • Accuthermo Technology Corp.
  • Welcome to Accuthermo Technology Corp.
  • A Total Thermal Solution Provider
home » product » Sensors & Transmitters » Transmitters » IST-D Microprocessor Based Programmable lsolated Signal Transmitter
Sensors & Transmitters » Transmitters
Product No. : IST-D
DIN rail mount
Programmable for various input signals and measuring range
Configurable without Power Connected
3750 Vrms isolation between input and output

Input:
Resistance thermometer (PT100)
Thermocouple (J,K,T,E,B,R,S,N,C)
Voltage/Current transmitter (mV/V/mA)

Output :
2-wire loop-power technology, 4 to 20 mA analog output.
Fault signal on sensor break presettable.
Part Number Builder
1
Part Number
 
The unit will come standard with PT100, -200~600°C, you can change the input Type/Rang using the free software "" with the configuration cable URC-1020, or you can contact us for non-standard Input/Rang setting.
 
Model No. Mount type
IST-D DIN Rail mount
 

The  IST-D is user configurable using our  software on your PC/notebook or our dedicated handheld programmer. The easy to use software is available free from your supplier or can be downloaded from our web site and should be used in conjunction with the URC-1020 USB cable which incorporates an interface converter necessary for communications.The IST-D transmitter can be configured without any loop or power connected making it easy to configure in the workshop before installation.
The unit will come standard with PT100, -200~600°C, you can change the input Type/Rang using the free software "" with the configuration cable URC-1020, or you can contact us for non-standard Input/Rang setting.
 

 
Specification
 
Input Thermocouple (T/C) : industry standard thermocouple types J, K, T, E, B, R, S, N, C (ITS-90).
Pt100: Excitation 180uA. 2 or 3 wire connection (ITS-90 α =0.00385).
Voltage: -60mVdc to 60mVdc or –10Vdc to 10Vdc.
Current : 0-24mAdc
Accuracy Refer to Table 1 Input Signal

A/D Resolution

16 bits

Input Sampling Rate

<200ms
Power supply DC 10 to 36V
Max. Load (V-10)/0.02 ( Ω )

Output Resolution

0.6 μ A(15 bits)

Output Response time

<200ms
Common Mode Rejection Ratio(CMRR) >80dB
Electromagnetic compatibility En 50081-2, En 50082-2
Galvanic isolation 3.75 KV. between input and output
Operating Temperature -40 to 85 ° C
Humidity 0 to 90% RH
Dimension 75mm(W)x87mm(H)x18.5mm(D)
 

 
Table 1 Input Signal
 
Input signal Maximum Range Accuracy
Thermocouple J -50 to 1000 ° C (-58 to 1832 ° F ) ± 1 ° C
Thermocouple K -50 to 1370 ° C (-58 to 2498 ° F ) ± 1 ° C
Thermocouple T -270 to 400 ° C (-454 to 752 ° F ) ± 1 ° C
Thermocouple E -50 to 700°C (-58 to 1292°F) ± 1 ° C
Thermocouple B 0 to 1750 ° C (32 to 3182 ° F ) ± 2 ° C (Note1)
Thermocouple R -50 to 1750 ° C (-58 to 3182 ° F ) ± 2 ° C
Thermocouple S -50 to 1750 ° C (-58 to 3182 ° F ) ± 2 ° C
Thermocouple N -50 to 1300 ° C (-58 to 2372 ° F ) ± 2 ° C
Thermocouple C -50 to 1800 ° C (-58 to 3272 ° F ) ± 2 ° C
Pt100 -200 to 600 ° C (-328 to 1112 ° F ) ± 0.2 ° C
mV -60mVto 60mV ± 0.01m V
Voltage (Note2) -10 to 10Vdc ± 1m V
Current (Note2) 0 to 24mAdc ± 10μA
 
Note 1 : Accuracy is not guaranteed between 0 and 400°C (0 and 752°F) for type B.
Note 2 : The internal DIP switch should be set.
 

 
Dimension
 
 

 