Welcome to Accuthermo Technology Corp.
-
A Total Thermal Solution Provider
Specifications
Interface:
|
Communication
|
RS-485, 2 wires
|
Baute rate (bps)
|
1200, 2400, 4800, 9600, 19.2K, 38.4K, 115.2K
|
LED indication
|
LED for power and communication
|
Analog input:
|
Input channels
|
6 differential + 2 single-ended or 8 differential
|
Resolution
|
16-bit / 12- bit
|
Accuracy |
+/- 0.1% |
Voltage range
|
+/-15mV ,+/-50 mV ,+/- 100mV, +/- 500mV,+/- 1V, +/-2.5V
|
Current measurement
|
+/-20mA (with external 125 ohms shunt resistor)
|
Thermocouple type
|
J, K, T, E, R, S, B, N, C
J: -210°C~760°C |
K: -270°C~1372°C |
T: -0270°C~400°C |
E: -0270°C~1000°C |
R: 0°C~ +1768°C |
S: 0°C~ +1768°C |
B: 0°C~ +1820°C |
N: -270°C~1300°C |
C: 0°C~2320°C |
|
Sample rate
|
10 samples/sec (16- bit) for normal mode
60 samples/sec (12-bit) for fast mode
|
Bandwidth
|
15.72Hz (normal mode), 78.6Hz (fast mode)
|
Zero drift
|
0.03 uV/C
|
Span drift
|
25 ppm/C
|
Accuracy
|
0.1% or better (normal mode)
|
CMRR
|
92 dB min. @50/60Hz
|
photo-isolation
|
3750Vrms
|
Isolation voltage
|
3000VDC
|
Power:
|
Power supply
|
+10VDC ~ +30VDC
|
Over voltage protection
|
+/- 35V
|
Power consumption
|
1.8W
|
Others:
|
Compatible
|
I-7018(F) / ADAM 4018
|
Manual
E8018_MANU_V11
Certification
|